High Purity NaI(Tl) Scintillator to Search for Dark Matter
Abstract
A high purity and large volume NaI(Tl) scintillator was developed to search for cosmic dark matter. The required densities of radioactive impurities (RIs) such as U-chain, Th-chain are less than a few ppt to establish high sensitivity to dark matter. The impurity of RIs were effectively reduced by selecting raw materials of crucible and by performing chemical reduction of lead ion in NaI raw powder. The impurity of 226Ra was reduced less than 100 µBq/kg in NaI(Tl) crystal. It should be remarked that the impurity of 210Pb, which is difficult to reduce, is effectively reduced by chemical processing of NaI raw powder down to less than 30 µBq/kg. The expected sensitivity to cosmic dark matter by using 250 kg of the high purity and large volume NaI(Tl) scintillator (PICO-LON; Pure Inorganic Crystal Observatory for LOw-background Neutr(al)ino) is 7 × 10-45 cm2 for 50 GeV/c2 WIMPs.
- Publication:
-
Proceedings of International Symposium on Radiation Detectors and Their Uses (ISRD2016
- Pub Date:
- 2016
- DOI:
- 10.7566/JPSCP.11.020003
- arXiv:
- arXiv:1605.04999
- Bibcode:
- 2016rdtu.confb0003F
- Keywords:
-
- Astrophysics - Instrumentation and Methods for Astrophysics;
- Physics - Instrumentation and Detectors
- E-Print:
- 6 pages, 2 Figures, Proceedings of International Symposium on Radiation Detectors and Their Uses (ISRD2016). Talk given on 19th Jan. 2016 by K.Fushimi. To be published in Proceedings will be published as JPS conference proceedings (2016)