Reducing current noise in cryogenic experiments by vacuum-insulated cables
Abstract
We measure the current noise of several cryogenic cables in a pulse tube based dilution refrigerator at frequencies between about 1 mHz and 50 kHz. We show that vibration-induced noise can be efficiently suppressed by using vacuum-insulated cables between room temperature and the 2nd pulse tube stage. A noise peak below 4 fA at the 1.4 Hz operation frequency of the pulse tube and a white noise density of 0.44 fA / √{ Hz } in the millihertz range are obtained.