Commissioning of a dual-phase xenon TPC at Nikhef
Abstract
A dual-phase xenon time-projection chamber was built at Nikhef in Amsterdam as a direct dark matter detection R&D facility. In this paper, the setup is presented and the first results from a calibration with a 22Na gamma-ray source are presented. The results show an average light yield of (5.6 ± 0.3) photoelectrons/keV (calculated to 122 keV and zero field) and an electron lifetime of (429±26) μs. The best energy resolution σE / E is (5.8 ± 0.2)% at an energy of 511 keV. This was achieved using a combination of the scintillation and the ionization signals. A photomultiplier tube gain calibration technique, based on the electroluminescence signals occurring from isolated electrons, is presented and its advantages and limitations are discussed.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- December 2016
- DOI:
- 10.1016/j.nima.2016.09.052
- arXiv:
- arXiv:1602.01974
- Bibcode:
- 2016NIMPA.840...87H
- Keywords:
-
- Xenon;
- TPC;
- Gain calibration;
- PMT;
- Single electron;
- Dark matter;
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- 14 pages, 17 figures