CHANTI: a fast and efficient charged particle veto detector for the NA62 experiment at CERN
Abstract
The design, construction and test of a charged particle detector made of scintillation counters read by Silicon Photomultipliers (SiPM) is described. The detector, which operates in vacuum and is used as a veto counter in the NA62 experiment at CERN, has a single channel time resolution of 1.14 ns, a spatial resolution of ~2.5 mm and an efficiency very close to 1 for penetrating charged particles.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- March 2016
- DOI:
- 10.1088/1748-0221/11/03/P03029
- arXiv:
- arXiv:1512.00244
- Bibcode:
- 2016JInst..11P3029A
- Keywords:
-
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- doi:10.1088/1748-0221/11/03/P03029