An apparatus for studying electrical breakdown in liquid helium at 0.4 K and testing electrode materials for the SNS nEDM experiment
Abstract
We have constructed an apparatus to study DC electrical breakdown in liquid helium at temperatures as low as 0.4 K and at pressures between the saturated vapor pressure and $\sim$600 torr. The apparatus can house a set of electrodes that are 12 cm in diameter with a gap of $1-2$ cm between them, and a potential up to $\pm 50$ kV can be applied to each electrode. Initial results demonstrated that it is possible to apply fields exceeding 100 kV/cm in a 1 cm gap between two electropolished stainless steel electrodes 12 cm in diameter for a wide range of pressures at 0.4 K. We also measured the current between two electrodes. Our initial results, $I<1$ pA at 45 kV, correspond to a lower bound on the effective volume resistivity of LHe of $\rho_V > 5\times10^{18}$ $\Omega\cdot$cm. This lower bound is 5 times larger than the bound previously measured. We report the design, construction, and operational experience of the apparatus, as well as initial results.
- Publication:
-
arXiv e-prints
- Pub Date:
- October 2015
- DOI:
- arXiv:
- arXiv:1510.06068
- Bibcode:
- 2015arXiv151006068I
- Keywords:
-
- Physics - Instrumentation and Detectors;
- Nuclear Experiment
- E-Print:
- 15 pages, 14 figures, accepted for publication in Review of Scientific Instruments