A large area plastic scintillation detector with 4-corner-readout
Abstract
A 760 mm × 760 mm × 30 mm plastic scintillation detector viewed by photomultiplier tubes (PMTs) from four corners has been developed, and the detector has been tested with cosmic rays and γ rays. A position-independent effective time T eff has been found, indicating this detector can be used as a TOF detector. The hit position can also be reconstructed by the time from the four corners. A TOF resolution of 236 ps and a position resolution of 48 mm have been achieved, and the detection efficiency has also been investigated.
Supported by National Natural Science Foundation of China (U1332207, 11405242)- Publication:
-
Chinese Physics C
- Pub Date:
- May 2016
- DOI:
- arXiv:
- arXiv:1508.05618
- Bibcode:
- 2016ChPhC..40e6001T
- Keywords:
-
- Physics - Instrumentation and Detectors;
- Nuclear Experiment
- E-Print:
- 5 pages, 9 figures, submitted to Chinese Physics C