Waveform timing performance of a 5 GS/s fast pulse sampling module with DRS4
Abstract
We first clarify timing issues of non-uniform sampling intervals regarding a 5 GS/s fast pulse sampling module with DRS4. A calibration strategy is proposed, and as a result, the waveform timing performance is improved to below 10 ps RMS. We then further evaluate waveform-timing performance of the module by comparing with a 10 GS/s oscilloscope in a setup with plastic scintillators and fast PMTs. Different waveform timing algorithms are employed for analysis, and the module shows comparable timing performance with that of the oscilloscope.
Supported by Knowledge Innovation Program of the Chinese Academy of Sciences (KJCX2-YW-N27), and National Natural Science Foundation of China (11175176)- Publication:
-
Chinese Physics C
- Pub Date:
- October 2015
- DOI:
- 10.1088/1674-1137/39/10/106101
- arXiv:
- arXiv:1501.00651
- Bibcode:
- 2015ChPhC..39j6101W
- Keywords:
-
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- doi:10.1088/1674-1137/39/10/106101