Note: Sub-Kelvin refrigeration with dry-coolers on a rotating system
Abstract
We developed a cryogenic system on a rotating table that achieves sub-Kelvin conditions. The cryogenic system consists of a helium sorption cooler and a pulse tube cooler in a cryostat mounted on a rotating table. Two rotary-joint connectors for electricity and helium gas circulation enable the coolers to be operated and maintained with ease. We performed cool-down tests under a condition of continuous rotation at 20 rpm. We obtained a temperature of 0.23 K with a holding time of more than 24 h, thus complying with catalog specifications. We monitored the system's performance for four weeks; two weeks with and without rotation. A few-percent difference in conditions was observed between these two states. Most applications can tolerate such a slight difference. The technology developed is useful for various scientific applications requiring sub-Kelvin conditions on rotating platforms.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- August 2014
- DOI:
- 10.1063/1.4891618
- arXiv:
- arXiv:1405.7576
- Bibcode:
- 2014RScI...85h6101O
- Keywords:
-
- Astrophysics - Instrumentation and Methods for Astrophysics;
- Physics - Instrumentation and Detectors
- E-Print:
- 3pages, 4 figures, to appear in Review of Scientific Instruments