Cross-calibration of the XMM-Newton EPIC pn and MOS on-axis effective areas using 2XMM sources
Abstract
Aims: We aim to examine the relative cross-calibration accuracy of the on-axis effective areas of the XMM-Newton EPIC pn and MOS instruments.
Methods: Spectra from a sample of 46 bright, high-count, non-piled-up, isolated on-axis point sources are stacked together, and model residuals are examined to characterize the EPIC MOS-to-pn inter-calibration.
Results: The MOS1-to-pn and MOS2-to-pn results are broadly very similar. The cameras show the closest agreement below 1 keV, with MOS excesses over pn of 0-2% (MOS1/pn) and 0-3% (MOS2/pn). Above 3 keV, the MOS/pn ratio is consistent with energy-independent (or only mildly increasing) excesses of 7-8% (MOS1/pn) and 5-8% (MOS2/pn). In addition, between 1-2 keV there is a "silicon bump" - an enhancement at a level of 2-4% (MOS1/pn) and 3-5% (MOS2/pn). Tests suggest that the methods employed here are stable and robust.
Conclusions: The results presented here provide the most accurate cross-calibration of the effective areas of the XMM-Newton EPIC pn and MOS instruments to date. They suggest areas of further research where causes of the MOS-to-pn differences might be found, and allow the potential for corrections and possible rectification of the EPIC cameras to be made in the future.
- Publication:
-
Astronomy and Astrophysics
- Pub Date:
- April 2014
- DOI:
- 10.1051/0004-6361/201423422
- arXiv:
- arXiv:1403.3555
- Bibcode:
- 2014A&A...564A..75R
- Keywords:
-
- instrumentation: detectors;
- instrumentation: miscellaneous;
- telescopes;
- X-rays: general;
- Astrophysics - Instrumentation and Methods for Astrophysics;
- Astrophysics - High Energy Astrophysical Phenomena
- E-Print:
- 8 Pages, 2 figures (3 panels), 1 table. Accepted for publication in A&