An atom trap trace analysis system for measuring krypton contamination in xenon dark matter detectors
Abstract
We have developed an atom trap trace analysis (ATTA) system to measure Kr in Xe at the part per trillion (ppt) level, a prerequisite for the sensitivity achievable with liquid xenon dark matter detectors beyond the current generation. Since Ar and Kr have similar laser cooling wavelengths, the apparatus has been tested with Ar to avoid contamination prior to measuring Xe samples. A radio-frequency plasma discharge generates a beam of metastable atoms which is optically collimated, slowed, and trapped using standard magneto-optical techniques. Based on the measured overall system efficiency of 1.2 × 10-8 (detection mode), we expect the ATTA system to reach the design goal sensitivity to ppt concentrations of Kr in Xe in <2 h.
- Publication:
-
Review of Scientific Instruments
- Pub Date:
- September 2013
- DOI:
- 10.1063/1.4821879
- arXiv:
- arXiv:1305.6510
- Bibcode:
- 2013RScI...84i3105A
- Keywords:
-
- astronomical instruments;
- cosmic ray apparatus;
- dark matter;
- krypton;
- sensitivity;
- 95.55.Vj;
- 95.35.+d;
- Neutrino muon pion and other elementary particle detectors;
- cosmic ray detectors;
- Dark matter;
- Physics - Atomic Physics;
- Astrophysics - Instrumentation and Methods for Astrophysics
- E-Print:
- 6 pages, 5 figures