A liquid xenon TPC for a medical imaging Compton telescope
Abstract
A new technique for medical imaging, “3γ imaging”, is studied by our group at SUBATECH for few years. A small liquid xenon time projection chamber prototype has been built in order to demonstrate the feasibility of this technique. With an ultra-low-noise front-end electronics, the energy deposit and resolution of 511 keV γ-ray as a function of drift electric field (E) is measured with high precision. 500 μm of z resolution is estimated by measuring the charge carriers drift velocity and time resolution.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- December 2012
- DOI:
- 10.1016/j.nima.2011.12.004
- arXiv:
- arXiv:1109.3586
- Bibcode:
- 2012NIMPA.695..125O
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- 8 pages, 8 figures, Proceeding of 6th international conference on New Developments In Photodetection, Lyon, 4-8 July 2011