Performance of prototypes for the ALICE electromagnetic calorimeter
Abstract
The performance of prototypes for the ALICE electromagnetic sampling calorimeter has been studied in test beam measurements at FNAL and CERN. A 4×4 array of final design modules showed an energy resolution of about 11%/√{E(GeV)}⊕1.7% with a uniformity of the response to electrons of 1% and a good linearity in the energy range from 10 to 100 GeV. The electromagnetic shower position resolution was found to be described by 1.5 mm⊕5.3 mm/√{E(GeV)}. For an electron identification efficiency of 90% a hadron rejection factor of >600 was obtained.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- March 2010
- DOI:
- 10.1016/j.nima.2009.12.061
- arXiv:
- arXiv:0912.2005
- Bibcode:
- 2010NIMPA.615....6A
- Keywords:
-
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- 10 pages, 10 figures