Trace Element Analysis of Potassium by Resonance Ionization Mass Spectrometry
Abstract
A simple resonance ionization mass spectrometer is built with a quadrupole mass spectrometer (QMS) and two laser diodes aiming at trace analysis of potassium. The overall detection efficiency is estimated comparing the yields of resonance ionization and electron-impact ionization in the same QMS. A clear discrimination of 39K, 40K, and 41K is demonstrated with a help of isotope shifts of the atomic levels.
- Publication:
-
Japanese Journal of Applied Physics
- Pub Date:
- July 2009
- DOI:
- 10.1143/JJAP.48.076505
- arXiv:
- arXiv:0808.4020
- Bibcode:
- 2009JaJAP..48g6505I
- Keywords:
-
- Physics - Atomic Physics
- E-Print:
- 14 pages, 9 figures, submitted to JJAP