Sensitivity and spatial resolution of square loop SQUID magnetometers
Abstract
We calculate the flux threading the pick-up coil of a square SQUID magnetometer in the presence of a current dipole source. The result reproduces that of a circle coil magnetometer calculated by Wikswo [J.P. Wikswo, AIP Conf. Proc. 44 (1978) 145] with only small differences. However it has a simpler form so that it is possible to derive from it closed form expressions for the current dipole sensitivity and the spatial resolution. The results are useful to assess the overall performance of the device and allow the comparison of magnetometers with different detection coil size.