Cavity BPM system tests for the ILC energy spectrometer
Abstract
The main physics programme of the International Linear Collider (ILC) requires a measurement of the beam energy at the interaction point with an accuracy of 10-4 or better. To achieve this goal a magnetic spectrometer using high resolution beam position monitors (BPMs) has been proposed. This paper reports on the cavity BPM system that was deployed to test this proposal. We demonstrate sub-micron resolution and micron level stability over 20 h for a 1 m long BPM triplet. We find micron-level stability over 1 h for 3 BPM stations distributed over a 30 m long baseline. The understanding of the behaviour and response of the BPMs gained from this work has allowed full spectrometer tests to be carried out.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- July 2008
- DOI:
- 10.1016/j.nima.2008.04.033
- arXiv:
- arXiv:0712.1350
- Bibcode:
- 2008NIMPA.592..201S
- Keywords:
-
- Physics - Instrumentation and Detectors
- E-Print:
- Paper submitted to Nuclear Instruments and Methods. 35 pages, 23 figures