Triple GEM performance in magnetic field
Abstract
Performance of triple GEM prototypes in strong magnetic field has been evaluated by means of a muon beam at the H4 line of the SPS test area at CERN. Data have been reconstructed and analyzed offline with two reconstruction methods: the charge centroid and the micro-Time-Projection-Chamber exploiting the charge and the time measurement respectively. Depending on the combination of the particle incident angle and magnetic filed, there's always one of the two algorithms achieving a spatial resolution of 100-120 μm.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- August 2019
- DOI:
- 10.1088/1748-0221/14/08/P08018
- arXiv:
- arXiv:1908.06253
- Bibcode:
- 2019JInst..14P8018A
- Keywords:
-
- Physics - Instrumentation and Detectors;
- High Energy Physics - Experiment
- E-Print:
- doi:10.1088/1748-0221/14/08/P08018