Radiation hardness of a p-channel notch CCD developed for the X-ray CCD camera onboard the XRISM satellite
Abstract
We report the radiation hardness of a p-channel CCD developed for the X-ray CCD camera onboard the XRISM satellite. This CCD has basically the same characteristics as the one used in the previous Hitomi satellite, but newly employs a notch structure of potential for signal charges by increasing the implant concentration in the channel . The new device was exposed up to approximately 7.9 × 1010 protons cm-2 at 100 MeV . The charge transfer inefficiency was estimated as a function of proton fluence with an 55Fe source. A device without the notch structure was also examined for comparison. The result shows that the notch device has a significantly higher radiation hardness than those without the notch structure including the device adopted for Hitomi. This proves that the new CCD is radiation tolerant for space applications with a sufficient margin.
- Publication:
-
Journal of Instrumentation
- Pub Date:
- April 2019
- DOI:
- arXiv:
- arXiv:1906.00171
- Bibcode:
- 2019JInst..14C4003K
- Keywords:
-
- Astrophysics - Instrumentation and Methods for Astrophysics;
- Physics - Instrumentation and Detectors
- E-Print:
- 10 pages, 6 figures, published in Journal of Instrumentation (JINST)