Randomized benchmarking of quantum gates implemented by electron spin resonance
Abstract
Spin systems controlled and probed by magnetic resonance have been valuable for testing the ideas of quantum control and quantum error correction. This paper introduces an X-band pulsed electron spin resonance spectrometer designed for high-fidelity coherent control of electron spins, including a loop-gap resonator for sub-millimeter sized samples with a control bandwidth ∼40 MHz. Universal control is achieved by a single-sideband upconversion technique with an I-Q modulator and a 1.2 GS/s arbitrary waveform generator. A single qubit randomized benchmarking protocol quantifies the average errors of Clifford gates implemented by simple Gaussian pulses, using a sample of gamma-irradiated quartz. Improvements in unitary gate fidelity are achieved through phase transient correction and hardware optimization. A preparation pulse sequence that selects spin packets in a narrowed distribution of static fields confirms that inhomogeneous dephasing (1 / T2∗) is the dominant source of gate error. The best average fidelity over the Clifford gates obtained here is 99.2 % , which serves as a benchmark to compare with other technologies.
- Publication:
-
Journal of Magnetic Resonance
- Pub Date:
- June 2016
- DOI:
- 10.1016/j.jmr.2016.04.010
- arXiv:
- arXiv:1510.04779
- Bibcode:
- 2016JMagR.267...68P
- Keywords:
-
- Electron spin resonance;
- Quantum information processing;
- Randomized benchmarking;
- Quantum Physics
- E-Print:
- 14 pages, 12 figures. Updated version. Accepted for publication in Journal of Magnetic Resonance