Measurements of rare isotopes of U and Th by MC-ICP-MS using a 1013 ohm resistor
Abstract
We have tested a 1013 ohm resistor on a Thermo-Scientific Neptune Plus, a multi-collector inductively-coupled plasma mass spectrometer (MC-ICP-MS), for the measurement of rare isotopes of uranium (U) and thorium (Th). In nature, the isotopic disequilibrium among U-series nuclides provides the potential to date materials and time processes over the last 700,000 years. Using gravimetric standards and a Minnesota stalagmite, we demonstrate the reproducibility of δ234U and 230Th dates with uncertainties at the 1-‰ to sub-‰ level (2σ), with relatively small samples. Compared to traditional secondary electron multiplier (SEM) techniques, measurement times decrease from > 1 hour to < 5 min for U and from tens of min to < 2 min for Th, with comparable or better precision. The characteristics of the new amplifier design and typical instrumental conditions allow for 234U and 230Th sample loads as small as 1-2 pg, a reduction in sample size close to an order of magnitude over cup measurements with 1011 ohm resistors. The main sources of error include the amplifier noise, uncertainty in the characterization of the tailing effect, and in some cases, counting statistics. Importantly, our overall characterization suggests that this new method forms the basis for future and further improvements on instrumental precision.