Peroxy Radical Measurements during PROPHET-AMOS 2016
Abstract
We present measurements of total peroxy radicals (HO2 + RO2) using the Ethane Chemical Amplifier (ECHAMP) technique during the PROPHET-AMOS project in Pellston, Michigan during July 2016. The C2H6/NO amplification chemistry occurred in FEP reaction chambers at the top of the PROPHET tower at a height of 34 m. The NO2 amplification product was transported through tubing to two cavity attenuated phase shift spectrometers (CAPS) housed inside the PROPHET laboratory. Two calibration sources were used: one based on water photolysis in the presence of isoprene and ozone actinometry, and another based on methyl iodide (CH3I) photolysis. The former was integrated into the inlet system, allowing for daily calibrations, whereas the latter was used twice during the campaign. Peak mixing ratios on warm, sunny days were approximately 40 ppt. Nighttime concentrations varied from below the instrumental detection limit to approximately 5 ppt. The measured peroxy radical concentrations will be compared to HO2 and HO2* mixing ratios measured by the Indiana University LIF-FAGE instrument.
- Publication:
-
AGU Fall Meeting Abstracts
- Pub Date:
- December 2016
- Bibcode:
- 2016AGUFM.A33A0188W
- Keywords:
-
- 0317 Chemical kinetic and photochemical properties;
- ATMOSPHERIC COMPOSITION AND STRUCTUREDE: 0365 Troposphere: composition and chemistry;
- ATMOSPHERIC COMPOSITION AND STRUCTUREDE: 0368 Troposphere: constituent transport and chemistry;
- ATMOSPHERIC COMPOSITION AND STRUCTUREDE: 3305 Climate change and variability;
- ATMOSPHERIC PROCESSES