Effective arrangement of separated transmit-only/receive-only RF coil for improvement of B1 homogeneity at 7 Tesla
Abstract
This article presents an effective arrangement with shifted transmit (Tx)-only and receive (Rx)-only (TORO) radiofrequency (RF) coils in a single-channel surface coil for improving the magnetic flux ( B 1) homogeneity in an ultra-high field (UHF) magnetic resonance imaging (MRI) scanner. The proposed new methodology for the coil arrangement using the shifted TORO RF coils was demonstrated for coils with 50-mm, 100-mm, and 150-mm-square surfaces and the results were compared to those for general Tx/Rx surface coils with the same dimensions. The computational analysis indicated that a homogeneous B1 field was achieved when the Rx-only coil was shifted in the two-dimensional xy-plane away from the Tx-only coils. Because the proposed coil configuration provides a unique opportunity for increasing the B 1 homogeneity, this feature is likely to increase the feasibility via new coil arrangements of UHF surface design and fabrication.
- Publication:
-
Journal of Korean Physical Society
- Pub Date:
- September 2014
- DOI:
- 10.3938/jkps.65.616
- Bibcode:
- 2014JKPS...65..616I
- Keywords:
-
- MRI (Magnetic Resonance Imaging);
- RF (Radiofrequency) coil;
- TORO (Transmit-only/receiveonly);
- Homogeneity;
- 7 Tesla