Using ACIS on the Chandra X-ray Observatory as a particle radiation monitor II
Abstract
The Advanced CCD Imaging Spectrometer is an instrument on the Chandra X-ray Observatory. CCDs are vulnerable to radiation damage, particularly by soft protons in the radiation belts and solar storms. The Chandra team has implemented procedures to protect ACIS during high-radiation events including autonomous protection triggered by an on-board radiation monitor. Elevated temperatures have reduced the effectiveness of the on-board monitor. The ACIS team has developed an algorithm which uses data from the CCDs themselves to detect periods of high radiation and a flight software patch to apply this algorithm is currently active on-board the instrument. In this paper, we explore the ACIS response to particle radiation through comparisons to a number of external measures of the radiation environment. We hope to better understand the efficiency of the algorithm as a function of the flux and spectrum of the particles and the time-profile of the radiation event.
- Publication:
-
Space Telescopes and Instrumentation 2012: Ultraviolet to Gamma Ray
- Pub Date:
- September 2012
- DOI:
- arXiv:
- arXiv:1207.3329
- Bibcode:
- 2012SPIE.8443E..11G
- Keywords:
-
- Astrophysics - Instrumentation and Methods for Astrophysics
- E-Print:
- 10 pages, 5 figures, to be published in Proc. SPIE 8443, "Space Telescopes and Instrumentation 2012: Ultraviolet to Gamma Ray"