Micro channel plate gains for Ta 10+-Ta 44+ ions, measured in the energy range from 3.7 keV/q up to 150.7 keV/q
Abstract
The procedure of calibration of a detector assembly consisting of a micro channel plate detector operated in a Chevron configuration is described. The analog particle gains of Ta ions with charge states ranging from q=10 + to 44 + and ion impact energies from 3.2 to 153.2 keV/ q have been measured. For the investigated range of ion impact energies we have stated that for tantalum charge states of 10 +-20 + the secondary ion-electron emission coefficient is dominated by the kinetic emission of electrons. Potential effects being visible for the charge states above 20 +.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- November 1999
- DOI:
- 10.1016/S0168-9002(99)00676-2
- Bibcode:
- 1999NIMPA.437..335M