Model-based pulse shape correction for CdTe detectors
Abstract
We present a systematic method to improve energy resolution of CdTe-detector systems with full control of the efficiency. Sampled pulses and multiple amplifier data are fitted by a model of the pulse shape including the deposited energy and the interaction point within the detector as parameters. We show the decisive improvements of spectral resolution and photo-peak efficiency that is obtained without distortion of spectral shape. The information concerning the interaction depth of individual events can be used to discriminate between beta particles and gamma quanta.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- February 1999
- DOI:
- 10.1016/S0168-9002(99)00546-X
- Bibcode:
- 1999NIMPA.434..399B