Optimization of Germanium Monochromators for Neutron Diffractometers
Abstract
A computer program TISTA has been employed to study the behavior of reactor-neutrons at the sample position of a neutron spectrometer diffracted from Ge(111), (220), and (311) monochromators. Our aim is to design a double axis neutron spectrometer and to determine the behavior of beam intensity and resolution at the sample position. The study will be helpful to design experiments with the existing triple axis neutron spectrometer at TRIGA Mark II research reactor, Dhaka, Bangladesh. The optimum values of crystal and instrument parameters have been determined through these calculations. The flux density of neutrons and the resolutions of a spectrometer at the sample position have been calculated as functions of beam collimation, zero-Bragg-angle deviation, crystal curvature, distance between sample and monochromator, crystal asymmetry, thickness, mosaic spread, crystal length, etc. The present results are compared with those of copper and silicon monochromators.
- Publication:
-
International Journal of Modern Physics E
- Pub Date:
- 1996
- DOI:
- 10.1142/S0218301396000062
- Bibcode:
- 1996IJMPE...5..131A