Calibration of multichannel analyzer of charge-exchange atoms
Abstract
The aim of the paper is calibration of units of a multichannel analyzer of charge-exchange neutrals intended for hot plasma diagnostics. A dependence is obtained between the microchannel-plate detector output current and fast H(sup 0) current at the analyzer input in the 5-80 keV range. Operation modes of a gas target and the amplification coefficient of the microchannel-plate detector are determined.
- Publication:
-
Unknown
- Pub Date:
- 1994
- Bibcode:
- 1994cmac.rept.....A
- Keywords:
-
- Calibrating;
- Electrostatic Probes;
- High Temperature Plasmas;
- Microchannel Plates;
- Plasma Diagnostics;
- Amplification;
- Energy Spectra;
- Ionization Chambers;
- Atomic and Molecular Physics