Recent developments of scientific CCDs have produced sensors that achieve ultra low read noise performance (less than 2 electrons rms) and near perfect charge transfer efficiency (0.9999996) without the addition of a fat-zero. This progress has now made it possible to achieve Fano-noise-limited performance in the soft X-ray where the detector's energy resolution is primarily limited by the statistical variation in the charge generated by the interacting X-ray photon. In this paper, Fano-noise-limited test data is presented for the different CCD types and a CCD derived estimate of the Fano factor is determined