Evaluation of industrial platinum resistance thermometers
Abstract
The calibration and stability of four surface temperature measuring industrial platinum resistance thermometers for use in the temperature range -120 C to 160 C was investigated. It was found that the calibration formulation of the International Practical Temperature Scale of 1968 provided the most accurate calibration. It was also found that all the resistance thermometers suffered from varying degrees of instability and hysteresis.