How to use electrodeless drift chambers in experiments at accelerators
Abstract
Several types of wide-gap electrodeless drift chambers, including those with dimensions 1×1 m 2, have been tested at accelerator beams. The chambers work with high efficiency (>99%), good spatial resolution ( σ = 0.2-0.4 mm) and good linearity at flux rates up to 2×10 5 particles -1 per wire in spill, which corresponds to 3×10 6 particles/s m 2 on colliders. It is shown that at the proper mode of operation beam intensity oscillations within the range of up to 4×10 5 particles/spill per wire do not affect the chamber efficiency.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- April 1987
- DOI:
- 10.1016/0168-9002(87)91217-4
- Bibcode:
- 1987NIMPA.255..493B