Accurate frequency measurements of noise-added Doppler signals by means of transient recorders and LDA counters using a laser diode simulator
Abstract
An experimental program is conducted to obtain detailed data on the SNR values desirable in laser Doppler anemometer (LDA) signals, as well as the sample points/cycle required in performing precise frequency measurements with transient recorders. It is noted that a systematic error can be introduced by incorrect hardware design. Comparisons between a commercially available LDA counter and a transient recorder demonstrate that a counter cannot be applied for uncertainties smaller than 0.1 percent.