Investigation of large format microchannel plate Z configurations
Abstract
The performance of triplet (Z) stacks of microchannel plates (MCPs) has been studied as a part of the instrument development for the Extreme Ultraviolet Explorer (EUVE) satellite mission. Relatively large MCPs with a 60-mm diameter and having a large 80:1 channel length to diameter (L:D) ratio were used in several configurations. The MCPs were used in the EUVE prototype imaging detector to provide more than 512 x 512 pixels with low image distortion (less than 1 percent). The gain and pulse height characteristics of the MCPs were examined, showing that both high gains (more than 2 x 10 to the 7th) and tight output pulse height distributions (less than 30 percent FWHM) may be achieved. Simple distribution techniques have also allowed low intrinsic background event rates (less than 0.15 events per sq cm/s) to be obtained. Variation of the quantum efficiency of the MCPs over the wavelength range 160-1216 A has been investigated for a range of angles of incidence. The effect of temperature variations on MCP operating characteristics has also been evaluated.
- Publication:
-
IEEE Transactions on Nuclear Science
- Pub Date:
- February 1985
- DOI:
- 10.1109/TNS.1985.4336871
- Bibcode:
- 1985ITNS...32..443S
- Keywords:
-
- Design Analysis;
- Extreme Ultraviolet Explorer Satellite;
- Microchannel Plates;
- Optimization;
- Radiation Detectors;
- Pulse Amplitude;
- Quantum Efficiency;
- Radiation Effects;
- Wavelengths;
- Spacecraft Instrumentation