Automatic ac/dc thermal voltage converter and ac voltage calibration system
Abstract
An automatic ac/dc difference calibration system is described which uses direct measurement of thermoelement emfs. In addition to ac/dc difference testing, the system can be used to measure some important characteristics of thermoelements, as well as to calibrate ac voltage calibrators and precision voltmeters. The system operates over a frequency range from 20 Hz to 100 kHz, covering the voltage range from 0.5 V to 1 kv. For all voltages the total measurement uncertainties expected (including the uncertainty of the specific reference thermal converters used) were 50 parts per million (ppm) at frequencies from 20 Hz to 20 kHz, inclusive, and 100 ppm at higher frequencies up to 100 kHz.
- Publication:
-
Final Report National Bureau of Standards
- Pub Date:
- November 1984
- Bibcode:
- 1984nbs..reptR....L
- Keywords:
-
- Automation;
- Calibrating;
- Systems Analysis;
- Systems Engineering;
- Voltage Converters (Ac To Ac);
- Computer Programs;
- Electric Potential;
- High Frequencies;
- Voltmeters;
- Instrumentation and Photography