Utilization of microchannel plates in X-ray crystallography
Abstract
An X-ray detector using a microchannel plate (MCP) was developed and utilized in X-ray crystallography. The tests show that this detector is more sensitive than a system scintillator-film, and that the main problem of scintillator and film, namely the possibly small distance to the sample for intensity reasons, does not occur with an MCP which detects the different X-ray quanta and has practically no dark current. This advantage is particularly interesting for X-ray crystallography on hollow-charge where the excited shock wave is susceptible to destroy a detector in the immediate vicinity. It is shown that image processing methods allow a better evaluation of the information available in the diagrams. The weak intensity line K beta 200, which is not visible with the eye on the X-ray image, is represented with the help of a computer.
- Publication:
-
Final Report Institut Franco-Allemand de Recherches
- Pub Date:
- December 1984
- Bibcode:
- 1984ifar.rept.....E
- Keywords:
-
- Crystallography;
- Microchannel Plates;
- X Ray Analysis;
- Image Processing;
- Product Development;
- X Ray Imagery;
- X Rays;
- Instrumentation and Photography