The Sacramento Peak fast microphotometer
Abstract
The Sacramento Peak Observatory Fast Microphotometer translates an optical system that includes a laser and photodiode detector across the film to scan the Y direction. A stepping motor moves the film gate in the X direction. This arrangement affords high positional accuracy, low noise (0.002 RMS density units), modest speed (5000 points/second), large dynamic range (4.5 density units), high stability (0.005 density units), and low scattered light. The Fast Microphotometer is interfaced to the host computer by a 6502 microprocessor.
- Publication:
-
Astronomical Microdensitometry Conference
- Pub Date:
- July 1984
- Bibcode:
- 1984asmd.conf..243A
- Keywords:
-
- Astronomical Photometry;
- Image Processing;
- Photometers;
- Accuracy;
- Computer Programs;
- Lasers;
- Low Pass Filters;
- Microprocessors;
- Numerical Control;
- Photodiodes;
- Scanning;
- Instrumentation and Photography