User's guide to calibration of analog electronic controllers in HVAC systems
Abstract
The calibration of analog electronic proportional controllers requires the knowledge of at least three characteristics of the controller: the action, the setpoint, and the throttling range. The calibration procedure follows these steps: measure the controller input(s); calculate the predicted controller output; measure the actual controller output, and adjust the controller as required. Using this procedure, controller performance is evaluated at the control point, or actual condition, rather than the setpoint, which is the idealized, desirable condition. This means that only those controllers which are actually out of calibration need to be adjusted and controllers that are in calibration need not have any settings or adjustments changed. If a controller cannot be calibrated or will not remain calibrated for a responsible period of time, it is defective and should be repaired or replaced.
- Publication:
-
NASA STI/Recon Technical Report N
- Pub Date:
- November 1984
- Bibcode:
- 1984STIN...8524229K
- Keywords:
-
- Analog Data;
- Calibrating;
- Controllers;
- Electronic Equipment;
- User Manuals (Computer Programs);
- Calibrating;
- Electrical Resistance;
- Ground Based Control;
- Input;
- Output;
- Systems Engineering;
- Throttling;
- Electronics and Electrical Engineering