An intercomparison of NEL and DHL water flow facilities using a twin orifice plate flowmeter assembly
Abstract
A 205 mm transfer standard orifice plate meter assembly, consisting of two orifice plates in series separated by a length of pipe containing a flow straightener, was calibrated in two water flow facilities. Results show that the agreement in the characteristics of such a differential pressure transfer standard package is within 0.17% over a 10:1 range from flow rates of approximately 8 to 80 l/sec. When the range over which the comparison was made was limited to that for which the calibration graphs gave straight lines, the agreement is 0.1% in 3 of the 4 calibrations (0.17% in the fourth).
- Publication:
-
NASA STI/Recon Technical Report N
- Pub Date:
- January 1983
- Bibcode:
- 1983STIN...8334271D
- Keywords:
-
- Flow Measurement;
- Flow Velocity;
- Flowmeters;
- Water Flow;
- Calibrating;
- Instrument Errors;
- Orifice Flow;
- Fluid Mechanics and Heat Transfer