Image plane detector for the dynamics explorer Fabry-Perot interferometer
Abstract
A compact twelve-channel photon-counting device based on existing Generation II imaging technology has been developed for use as the image-plane detector of the Dynamics Explorer Fabry-Perot interferometer. The device has an S-20 photocathode, three-microchannel plate electron multiplication stages, and an equal-area concentric-ring segmented anode whose geometry mimics that of the interference ring pattern produced by a plane etalon. The twelve channels sample equal and contiguous intervals in the spectrum. The purpose of the delopment has been to utilize the signal multiplex advantage of a multichannel detector in the measurement of Doppler shifts and line-broadening effects for naturally occurring atmospheric emission features of low intensity. The design, testing, calibration, and flight performance of the novel detector system are presented. In addition, measured emission line profiles at high resolution from the satellite instrument are presented to illustrate the operation of the device.
- Publication:
-
Applied Optics
- Pub Date:
- November 1983
- DOI:
- 10.1364/AO.22.003503
- Bibcode:
- 1983ApOpt..22.3503K
- Keywords:
-
- Dynamics Explorer 1 Satellite;
- Fabry-Perot Interferometers;
- Ionospheric Sounding;
- Photometers;
- Satellite-Borne Instruments;
- Emission Spectra;
- Imaging Techniques;
- Photocathodes;
- Thermosphere;
- Spacecraft Instrumentation;
- PHOTON COUNTING;
- ATMOSPHERES;
- INTERFEROMETRY;
- FABRY PEROT