Potential application of a single-particle aerosol spectrometer for monitoring aerosol size at the DOE filter test facilities
Abstract
A commercial single particle aerosol spectrometer, the LAS-X, was used to demonstrate that the test aerosol produced by each of the Q107 high efficiency particulate air filter test systems was polydisperse with a geometric standard deviation greater than 1.35, and a geometric mean diameter less than 0.2 microns. Use of the LAS-X for monitoring aerosol size was evaluated. The LAS-X is relatively insensitive to changes in aerosol concentration and refractive index over the range of interest for filter testing. It appears that the LAS-X would provide a better instrument for monitoring aerosol size, and would permit the test operator to adjust the test aerosol to approximate the size characteristics recommended for this test.
- Publication:
-
Presented at 17th DOE Nucl. Air Cleaning Conf
- Pub Date:
- 1982
- Bibcode:
- 1982nac..conf.....S
- Keywords:
-
- Aerosols;
- Air Filters;
- Particle Size Distribution;
- Spectroscopic Analysis;
- Mean;
- Quality Control;
- Refractivity;
- Standard Deviation;
- Instrumentation and Photography