Intercomparison of aerosol instruments: Number concentration
Abstract
An intercomparison of aerosol measuring instruments focused on five instruments: the Pollak and TSI condensation nucleus counters; the Active Scattering Aerosol Spectrometer (ASAS-X); and two aerosol electrometers. Test aerosols of sodium chloride and ammonium fluorescein generated by nebulization/electrostatic classification were used to obtain 195 lines of comparison data. Concentrations measured by the ASAS-X and the TSI aerosol electrometer averaged respectively 1.388 and 1.581 times that measured by the Pollak. These ratios were very stable during the week and there was little effect of particle size or material. Most other comparisons were equally stable. However, a review of past work led to the disturbing conclusion that these ratios may change from year to year, or from season to season. A filter sample was taken from microscopy, concurrent with readings from the ASAS-X and the TSI condensation nucleus counters. In this sample, the two instruments differed by 20%. Within its 20% uncertainty, the filter result matched both the TSI and ASAS-X readings.
- Publication:
-
NASA STI/Recon Technical Report N
- Pub Date:
- May 1982
- Bibcode:
- 1982STIN...8334297K
- Keywords:
-
- Aerosols;
- Atmospheric Composition;
- Concentration (Composition);
- Condensing;
- Electrometers;
- Scatterometers;
- Spectrometers;
- Ammonium Compounds;
- Comparison;
- Ratios;
- Sodium Chlorides;
- Instrumentation and Photography