Reproducibility of an automated laser Raman system for light gas analysis
Abstract
The experimental design, method of data collection, and subsequent analysis of an experimental laser Raman system are described. Results describe the feasibility of application of this system for light gas analysis. It is shown that the 1-sigma standard deviation reproducibilities of 0.03% can be obtained for major isotopic components (97%) and 0.9% reproducibility for a 12.3% component. Gases present at 0.49% can be analyzed with better than 4% reproducibility.
- Publication:
-
NASA STI/Recon Technical Report N
- Pub Date:
- May 1982
- Bibcode:
- 1982STIN...8324841S
- Keywords:
-
- Gas Analysis;
- Hydrogen Isotopes;
- Photons;
- Laser Spectroscopy;
- Photomultiplier Tubes;
- Raman Lasers;
- Lasers and Masers