ECC ozonesonde performance at high altitudes: Pump efficiency
Abstract
The ECC ozonesonde sampling behavior was examined at pressures ranging from 60 to 6 hPa with the objective of evaluating uncertainties in high altitude ozone data caused by variations in pumping efficiency. The averaged pump efficiency correction curve for a 43 sample set of 3A type ECC pumps showed a 2-3% bias from the curve provided by the manufacturer. In addition, random pump to pump variations (2 sigma) + or - were 5% at 6 hPa. These values probably represent minimum errors since the pumps were from the same production batch. A seven sample set of the newer 4A type ECC pumps was examined, with similar findings as for the 3A types.
- Publication:
-
ECC ozonesonde performance at high altitudes: Pump efficiency
- Pub Date:
- May 1981
- Bibcode:
- 1981eoph.rept.....T
- Keywords:
-
- Balloon-Borne Instruments;
- Compressor Efficiency;
- Ozonometry;
- Performance Tests;
- Pumps;
- Sondes;
- Atmospheric Sounding;
- High Altitude;
- Pressure Distribution;
- Vertical Distribution;
- Geophysics