Modified MEE Industries static thermal gradient diffusion cloud chamber
Abstract
The experiments performed at the Workshop were undertaken to confirm the calculated sample volume, determine the usable range of supersaturation, and minimum detectable size. Comparison of absolute CCN concentrations with other state-of-the-art continuous flow diffusion chambers at supersaturations near 1 percent indicated that the volume used produced CCN concentrations well within the range of CCN concentrations determined at the workshop. This agreement is interpreted to mean the sample volume was correct. Direct measurements of the beam geometry done in the laboratory indicated a factor of 4-5 error. This error is apparently due to the larger apparent visible beam diameter versus the actual usable beam diameter given the droplet illumination, chamber optical geometry, microscope optics and film characteristics.
- Publication:
-
The Third International Cloud Condensation Nuclei Workshop
- Pub Date:
- November 1981
- Bibcode:
- 1981clcn.work...44B
- Keywords:
-
- Cloud Chambers;
- Condensing;
- Ice Nuclei;
- Nucleation;
- Particle Density (Concentration);
- Temperature Gradients;
- Thermal Diffusion;
- Aerosols;
- Cloud Physics;
- Drop Size;
- Optical Measuring Instruments;
- Precipitation Particle Measurement;
- Instrumentation and Photography