Abstract
The paper presents an analysis and the results of calculation of the systematic error for the 6-15-micron spectral region - caused by a difference between the zero of the measurement scale of the recorder and the radiation zero - for measurements made with a single-beam spectroradiometer having a blackened Duralumin modulator. Idealized sources with a circular brightness indicatrix are discussed. It is shown that the systematic error can be of substantial value.
- Publication:
-
Optiko Mekhanicheskaia Promyshlennost
- Pub Date:
- September 1978
- Bibcode:
- 1978OpMeP..45...60U
- Keywords:
-
- Error Analysis;
- Infrared Radiometers;
- Optical Measurement;
- Spectroradiometers;
- Black Body Radiation;
- Signal Analysis;
- Instrumentation and Photography