Multichannel HCN interferometer for electron density profile measurements of tokamak plasmas
Abstract
A multichannel interferometer designed for Tokamak plasma-density measurements is described. The phase shift experienced by the probing beam is transferred to a low-frequency signal obtained by Doppler shifting the frequency of the reference beam. This ensures a sensitivity of 0.01 fringe. The source is a CW HCN waveguide laser with 150-mW output power. Eight channels are in operation at a certain Tokamak facility. The variation of line density as a function of time for each channel is deduced from the phase-shift measurement by an on-line computer.