Calibration of electron-energy spectrometers
Abstract
Relative photoionization cross-sections of strong peaks in the He(I) photoelectron spectra of N2, CO, CO2, and O2 are tabulated. These data have been measured with an electron energy analyzer whose relative luminosity has been calibrated to an accuracy of + or - 5%. Thus, the tables will be useful for calibrating the transmission of other analyzers for electron energies below 9 eV. Correction for angular distribution effects is discussed.