Development of a dynamic pressure calibration technique
Abstract
The report deals with work continuing on the development of a method of producing sinusoidally varying pressures of at least 34 kPa zero-to-peak with amplitude variations within plus or minus 5% up to 2 kHz for the dynamic calibration of pressure transducers. Sinusoidally varying pressures of 34 kPa zero-to-peak were produced between 40 Hz and 750 Hz by vibrating a 10-cm column of a dimethyl siloxane liquid at 36gn zero-to-peak. Damping of the liquid column was accomplished by packing the fixture tube with a number of smaller diameter tubes.
- Publication:
-
Progress Report
- Pub Date:
- June 1975
- Bibcode:
- 1975nbs..reptQ....V
- Keywords:
-
- Calibrating;
- Dynamic Pressure;
- Pressure Sensors;
- Columns (Process Engineering);
- Electric Choppers;
- Liquids;
- Siloxanes;
- Instrumentation and Photography