DALI2: A NaI(Tl) detector array for measurements of γ rays from fast nuclei
Abstract
A NaI(Tl) detector array called DALI2 (Detector Array for Low Intensity radiation 2) has been constructed for in-beam γ-ray spectroscopy experiments with fast radioactive isotope (RI) beams. It consists typically of 186 NaI(Tl) scintillators covering polar angles from ~15° to ~160° with an average angular resolution of 6° in full width at half maximum. Its high granularity (good angular resolution) enables Doppler-shift corrections that result in, for example, 10% energy resolution and 20% full-energy photopeak efficiency for 1-MeV γ rays emitted from fast-moving nuclei (velocities of v / c ≃ 0.6). DALI2 has been employed successfully in numerous experiments using fast RI beams with velocities of v/c=0.3-0.6 provided by the RIKEN RI Beam Factory.
- Publication:
-
Nuclear Instruments and Methods in Physics Research A
- Pub Date:
- November 2014
- DOI:
- arXiv:
- arXiv:1403.5349
- Bibcode:
- 2014NIMPA.763..596T
- Keywords:
-
- Scintillation detector;
- In-beam γ-ray spectroscopy;
- Radioactive beam experiment;
- Nuclear Experiment
- E-Print:
- 10 pages, 8 figures